Short Name:
A400c
Instrument Type:
NMR & EPR
Location:
KCL 3
Date of Acquisition:
October 2013
Description:
2 channel system with AS7620 96-sample changer. OneNMR probe with auto-tune-and-match installed.
Primarily used for short, routine NMR experiments where data is needed quickly.
Guides and Standard Operating Procedures
Best experiments:
- PROTON
- COSY
- Other nuclei: F19 (no H1 decoupling), P31, B11, (others by request)
Acceptable experiments:
- C13CPD (night time)
- 2D HSQC and HMBC experiments (night time)
- 1D/2D NOE experiments (night time)
Not available:
- Variable Temperature experiments
- Kinetics experiments
- Simultaneous F19 and H1 experiments
Time Required for NMR Experiments on A400c:
Mass | Concentration | Proton | Carbon | gCOSY | gHSQC | gHMBC |
0.03 mg | 0.1 mM | 10 h 40 min | 35389 hours | 1785 hours | 1886 hours | 7129 hours |
0.06 mg | 0.2 mM | 2 h 40 min | 10516 hours | 46 h 28 min | 474 h 12 min | 1785 hours |
0.15 mg | 0.5 mM | 0 h 26 min | 1415 h 33 min | 6 h 10 min | 75 h 01 min | 284 h 44 min |
0.3 mg | 1 mM | 0 h 06 min | 354 h 01 min | 1 h 40 min | 18 h 53 min | 71 h 19 min |
0.6 mg | 2 mM | 0 h 01 min | 88 h 31 min | 0 h 26 min | 4 h 42 min | 17 h 51 min |
1.5 mg | 5 mM | < 1 min | 14 h 09 min | 0 h 03 min | 0 h 45 min | 2 h 51 min |
3 mg | 10 mM | < 1 min | 3 h 32 min | 0 h 03 min | 0 h 16 min | 1 h 07 min |
6 mg | 20 mM | < 1 min | 0 h 53 min | 0 h 03 min | 0 h 08 min | 0 h 17 min |
12 mg | 40 mM | < 1 min | 0 h 13 min | 0 h 03 min | 0 h 08 min | 0 h 17 min |
15 mg | 50 mM | < 1 min | 0 h 08 min | 0 h 03 min | 0 h 08 min | 0 h 17 min |
> 30 mg | 100+ mM | < 1 min | 0 h 04 min | 0 h 03 min | 0 h 08 min | 0 h 17 min |
All estimates based on a 500 MW sample in 600 uL solvent volume using a 5 mm diameter NMR tube