Diffusion EXP on Bruker

Diffusion EXP on Bruker

1.  How to calibrate the gradient strength?

Book Experiment 11.1.

Check gradient strength and the file: (/NMRHOME/data/user/nmr/dosy2ddatafile/difflist.)

gradient strength  is saved in the file: /NMRHOME/conf/instr/gradient_calib

It is 5.35 G/mm for b400b.

Use Shigemi tube, 1 cm height sample, to run diffusion EXP.

Following Peter’s script. (comments:  Including the process in TOPSPIN).

Process the 2d dosy date with T1/T2 analysis on XWINNMR..

Following manual ‘dosy_diffusion_4XWINNMR3’

2. How to make diffusion exp easier:

EXP 1 ‘rpar PROTON all’, ‘getprosol’.

EXP 2 PROTON, record

SW o1p AQ for EXP 3.

EXP 3 ‘rpar DIFFSION all’, ‘getprosol’

Modify SW o1p AQ.

‘rga’ ‘zg’ ‘efp’

Copy EXP 3 to EXP4, and change ‘gpz6=95’.

(No ‘rga’) ‘zg’ ‘efp’

Compare EXP3 and EXP4.  Adjust d20 or p30 if necessary.

After changing p30, do not use ‘getprosol’.  Otherwise, p30 will change back to 1800(default value).

Continue to EXP 5 as described in the manual.